DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 8 - Replacement Procedures 8 - 79
8-9-16 Restore Standard Printer IP and Port Settings - Software Version v2.x (cont’d)
9.) Verify Printer Name or IP Address as recorded in Table 8-18 on page 8-67.
10.)Select Raw.
11.)Verify Port Number as recorded in Table 8-18 on page 8-67.
12.)Select ok.
13.)Close the rest of the Windows by selecting ok.
8-9-16-1 DC Offset Calibration
1.) Disconnect all probes.
2.) From the upper section of the window, select Test.
3.) Then select Go to System Test.
4.) From the Module selection block, select DC Offs. Cal.
5.) Reboot the System when the DC Offset Calibration has been completed.
8-9-16-2 Verify System Software Version
1.) Press CONFIG
(F2) and log on as adm, see 4-2-4 "Log On to the System as ‘ADM’" on page 4-12.
2.) Select About tab.
3.) Verify that the SW versions are the same as what were recorded in Table 8-20 "Record Software
and hardware versions" on page 8-69.
8-9-17 Verification - Functional Checks
Please perform the tests specified in Section 8-20 "Verification - Functional Checks" on page 8-105.
Figure 8-77 Configure Port
Raw
Port Number
ok
Printer Name or IP Address