GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 5 - Components and Functions (Theory) 5 - 117
Section 5-8
External I/O
5-8-1 General Description, External I/O
The External I/O is the interface between the scanner and all external devices.
Several signal are fed from the Internal I/O module to the External I/O module. The External I/O module
contains data buffers and a galvanic isolation section.
Different types of signals are fed to connectors accessible on the External I/O:
R, G, B and Sync are provided for external RGB monitors or printers.
Composite video, black and white video and S-VHS video are buffered and fed to output connectors
for external monitors and printers.
Two serial RS-232 ports are provided for external communication with e.g. the System CPU. One
of the RS-232C ports are located inside the scanner, and was intended for use by a modem, the
other is located on the EIO rear panel. These serial ports have never been used on Vivid 7.
An Ethernet signal from the BEP is fed to an Ethernet connector. It is used for connectivity, including
DICOM Network.