EasyManua.ls Logo

GE Vivid 7 - Fep2

GE Vivid 7
858 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 5 - Components and Functions (Theory) 5 - 7
5-3-3 FEP2
In units with FEP2, a new card, the Radio Frequency Interface (RFI) card replaces four of the cards
used in FEP1.
The cards used for FEP2 units are listed in Table 5-3 "Front-End Processor Cards used in FEP2"
on page 5-7.
Table 5-3 Front-End Processor Cards used in FEP2
SHORT
NAME COMPLETE NAME COMMENT
RLY
RELAY BOARD
TX
TRANSMITTER BOARD
2X TO BE USED PER UNIT FOR 3D OPTION
( ONLY, NOT FOR PRO))
RX
RECEIVER BOARD
BF
BEAM FORMER BOARD
2X USED PER UNIT
RFI
RADIO FREQUENCY INTERFACE BOARD
XDBUS
TRANSDUCER BUS BOARD
2X USED PER UNIT

Table of Contents