DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 4 - Functional Checks 4 - 69
4-3-18 Cineloop Check
4-3-18-1 Introduction
A cineloop is a sequence of images recorded over a certain time frame. When using ECG the time frame
can be adjusted to cover one or more heart cycles. When frozen the System automatically displays the
cineloop boundary markers on either side of the last detected heart cycle (Figure 4-36).
4-3-18-2 Preparation
Use a phantom when doing these tests.
1.) Connect one of the probes, to the scanner’s left-most probe connector.
-See 3-6-5 "Probe Connection" on page 3-21 for info about connecting the probes
-See Section 9-26 "Probes" on page 9-80 for available probes
2.) Turn ON the scanner
-See 4-2-2 "Power ON/ Boot UP" on page 4-3 for info about booting the scanner.
The 2D Mode window is displayed (default mode).
3.) Press Freeze to freeze the image.
4.) Press the Mode Dependant button Cineloop.
Figure 4-36 The Cineloop controls display
Left marker
Right marker
Current frame
Cine speed
(replay)
Loop display
ECG