EasyManua.ls Logo

GE Vivid 7 - Configuration; VIVID 7 Configuration

GE Vivid 7
858 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
3 - 28 Section 3-7 - . Configuration
Section 3-7.
Configuration
3-7-1 Vivid 7 Configuration
3-7-1-1 Select System Settings Screen
1.) Press CONFIG
(F2) and log on as ADM, see 4-2-4 "Log On to the System as ‘ADM’" on page 4-12.
NOTE: For software version v5.x and earlier the default password for ADM (or: adm) was left blank.
Starting with software version v6.x, the password is: ulsadm
2.) Select System and then select Settings, if needed.
3-7-1-2 Enter Location
Figure 3-18 Hospital and Department Name
Table 3-11 Location Name
STEP TASK EXPECTED RESULT(S)
1
Select the Hospital field, see Figure 3-18, and type the name of
the hospital (max 64 characters).
The 24 first characters of this name are displayed on the scanning
screen’s title bar (after restart)
All 64 are displayed on the image properties on saved images (after
restart).
2
Select in the Department field, see Figure 3-18, and type the
name of the department (max 64 characters).
This name will be displayed on the image properties on saved images as
soon as the unit has been restarted.
HOSPITAL NAME
DEPARTMENT