DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 4 - Functional Checks 4 - 27
4-2-13 Moving and Transporting the Vivid 7 (cont’d)
Place the probes in their carrying case.
Lock Operator Panel (Control Console) in place.
The Operator Panel (Control Console) must be locked into the center (x/y) position as indicated on
the label on the Front Handle, before transporting the system.
Eject any Magneto Optical disk, CD or DVD from their drive.
Secure the unit in an upright position.
Lock the casters (wheels) (brake)
DO NOT use the Operator Panel (Control Console) as an anchor point.
NOTE: Special care should be taken when transporting the unit in a vehicle:
Please refer to the Vivid 7 Unpacking/Packing Procedure, Direction Number FC050172 - Section 4-3
“Loading and Transporting the Unit in a Van”, if you need to transport the Vivid 7 by vehicle, and don’t
have the transportation cage available.
Figure 4-24 Label on Front Handle
OR
OR