EasyManua.ls Logo

GE Vivid 7 - Front-End Processor (FEP); General Description; Fep1

GE Vivid 7
858 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
5 - 6 Section 5-3 - a Front-End Processor (FEP)
Section 5-3a
Front-End Processor (FEP)
5-3-1 General Description
The electronics in the Front-End Card Cage is also called the Front-End Processor (FEP).
Two versions of the Front-End Card Cage are in use;
Version 1 (FEP1) was used from introduction of Vivid 7 until 2nd half of 2003
Version 2 (FEP2) was introduced 2nd half of 2003
5-3-2 FEP1
FEP1 has eight different types of electronics cards, totally 10 cards. The cards are listed in Table 5-2.
NOTICE
Be careful so you don’t insert a card in the wrong position in the card rack. If the power is turned on with
a card placed in the wrong position, the Vivid 7 will be destroyed.
Table 5-2 Front-End Processor Cards used in FEP1
SHORT
NAME COMPLETE NAME COMMENT
RLY
RELAY BOARD
TX
TRANSMITTER BOARD
RX
RECEIVER BOARD
BF
BEAM FORMER BOARD
2X USED PER UNIT
FEC
FRONT-END CONTROLLER BOARD
RFT
RADIO FREQUENCY & TISSUE BOARD
SDP
SPECTRUM DOPPLER PROCESSOR BOARD
IMP
IMAGE PORT
SOMETIMES CALLED “IMPORT”
XDBUS
TRANSDUCER BUS BOARD
2X USED PER UNIT

Table of Contents