EasyManua.ls Logo

GE Vivid 7 - Receiver Board, RX-128

GE Vivid 7
858 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 5 - Components and Functions (Theory) 5 - 31
5-3-14 Receiver Board, RX-128 (cont’d)
5-3-14-5 DC Voltages
5-3-14-6 Input Signals
5-3-14-7 Output Signals
5-3-14-8 Fuses
None
5-3-14-9 Jumpers and Dip-switches
None
Table 5-20 DC Input Voltages, Plug P3 & P4
INPUT DESCRIPTION CONNECTED TO
+ 5 Va
DC output voltage, distributed via Motherboard
(Backplane).
The “a” indicates that this voltage is used for analog circuits
DC Power Supply via Backplane
- 5 Va
DC output voltage, distributed via Motherboard
(Backplane).
The “a” indicates that this voltage is used for analog circuits.
DC Power Supply via Backplane
+15 V DC output voltage, distributed via Motherboard (Backplane) DC Power Supply via Backplane
- 15 V
DC output voltages, distributed via Motherboard
(Backplane)
DC Power Supply via Backplane
GND Distributed via Motherboard (Backplane) DC Power Supply via Backplane
Table 5-21 Input Signals
INPUT DESCRIPTION CONNECTION FROM
XD1 - XD128
Signals containing the received back-scattered echoes
from the active probe.
Active Probe > Relay Board > XD Bus
Boards > RX128 board
PEDOF
Signals containing the received back-scattered echoes
from PEDOF probe
Pedof probe > Relay Board > XD Bus
Boards > RX128 board
ATGCVP/ATGCVN TGC Control Voltages
FEC Board/RFI Board > FE Bus >
RX128 board
Test SIG Analog test signal
FEC Board/RFI Board > FE Bus >
RX128 board
Table 5-22 Output Signals
OUTPUT DESCRIPTION CONNECTION TO
RX1 - RX128 128 RX signals (echoes) Beam Former Boards via Backplane

Table of Contents