EasyManua.ls Logo

GE Vivid 7 - Front-End Alignment Procedure

GE Vivid 7
858 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 6 - Service Adjustments 6 - 7
6-7-3 Front-End Alignment Procedure
1.) Disconnect all connected probes.
2.) Turn on the Vivid 7.
3.) Press CONFIG
(F2) and log on as adm, see 4-2-4 "Log On to the System as ‘ADM’" on page 4-12.
4.) Select System
5.) Select DC Offs. Cal to start the DC Offset calibration of the Beam Formers (Front-End Alignment).
6.) Wait for the calibration to take place.
Front-End Alignment Description
- Each A/D converter will get 0V input (by setting ATGC to minimum).
- One by one channel will be digitized and the result put in the IQ buffer on the RFT board.
- The BEP will then read the IQ buffer and computes a value per channel. These computed
values are compared to an upper/lower limit.
- If all read values are ok, the message PASSED will appear. If values exceeding the limits are
read, the system will prompt you with FAILED and a message will tell you which BF board is
bad.
- The resulting offsets for each channel are stored in a file on the Back-End Processor’s hard
disk.
- The values can be plotted in a diagram Show graph).
7.) When finished, select Show Graph.
Wavy line signifies system has been calibrated. See Figure 6-3 "DC Calibration of the Beam Former
AD-converters Passed" on page 6-8.
A Straight line signifies that the system has not been calibrated and can be the cause of image
quality issues.
Figure 6-2 Calibration Dialog
Channel Control
DC Offs. Cal
Cal. Reset
Show graph

Table of Contents

Related product manuals