GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
8 - 18 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x
8-7-7 Backup - Software Version v7.x/v6.x/v5.x/v4.x (cont’d)
10.)Select Start backup. The following situations may occur:
- The system is checking that the removable media is inserted. If not, a dialogue window is
displayed prompting the user to insert a media.
- Insert the media and select OK.
- The system is checking if the media needs to be formatted. If yes, the media is automatically
formatted. An Information window is displayed showing the media label.
- Record the label and select OK.
- The system is checking if there is already a backup or a Disk management copy on the media.
If the following error message is displayed, the disk is ejected and the user is asked to use a
new media that does not contain any backup or Disk management data.
Figure 8-12 The information window
Figure 8-13 The replace current media window
Figure 8-14 The backup progress windows