DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7 - 1
Chapter 7
Diagnostics/Troubleshooting
Section 7-1
Overview
7-1-1 Purpose of this Chapter
This section describes how to setup and run the tools and software that help maintain image quality and
system operation. Very basic host, system and board level diagnostics are run whenever power is
applied. Some Service Tools may be run at the application level.
NOTE: Diagnostics must be run from start-up.
NOTE: After Diagnostics are performed/run, the system must be rebooted. If not done, this could cause
subsequent false failures due to the machine being left in an unknown state.
7-1-2 Contents in this Chapter
Table 7-1 Contents in Chapter 7
Section Description Page Number
7-1 Overview 7-1
7-2 Service Tools Overview 7-2
7-3 Frequently Asked Questions - FAQ 7-3
7-4 Service Safety Considerations 7-4
7-5 Gathering Troubleshooting Data 7-5
7-6 Diagnostics Tools 7-7
7-7 Common Diagnostics 7-48
7-8 Acquisition Diagnostics 7-63