DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 8 - Replacement Procedures 8 - 89
Section 8-13
TX Power Replacement Procedure
8-13-1 Manpower
One person, 30 minutes
8-13-2 Tools
Phillips screwdriver 2
4 mm blade screwdriver
8-13-3 Preparations
1.) Shut down the system, see 4-2-3 "Power Shut Down" on page 4-7.
2.) Remove Right Side Cover, see "Side Covers Removal Procedure" on page 8-4.
3.) Remove the Front-End Door.
a.) Turn seven (7) quarter-turn screws to loosen Front-End door.
b.) Holding the door, lift it out of its retainers along the lower edge of Front-End.
4.) Disconnect PCI cable from IMP2 Board shown in Figure 8-84.
.
DANGER
Dangerous voltages are present in this equipment. Completely power off and
shut down the scanner before servicing.
Figure 8-84 DC Power, TX Power and Cabling.
HV power cable
DC power cable
PCI cable
IMP2 Board
TX Power
Retaining clasp
DC Power