EasyManua.ls Logo

GE Vivid 7 - Input Signals

GE Vivid 7
858 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 5 - Components and Functions (Theory) 5 - 119
5-8-3 Input Signals
Figure 5-75 EIO - Overview
Table 5-79 Input Signals
Input Description Connection from/via/to
Footswitch
Connector
Footswitch is a mechanical switch that connect a signal to
ground.
Internal level is TTL (0-3.3V)
Footswitch > External I/O > Internal I/O
> PC2IO (BEP)
DC Voltages from BEP
+5 VDC
Used for Optical Couplers on RS232 signals
PC2IO (BEP) > Internal I/O > External I/
O
+12 VDC
DC Voltages from Card Cage
+5 VDC
Used for Optical Couplers on Footswitch Card Cage > Internal I/O > External I/O
+15 VDC
-5 VDC
analog
+6 VDC
analog

Table of Contents