GE Vivid 7 - Introduction; Preparations before Software Re-Installation

GE Vivid 7
858 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
8 - 12 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x
Section 8-7
Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x
8-7-1 Introduction
8-7-2 Preparations Before Software Re-installation
Export Patient Images from the internal hard disk to removable removable media.
Make backup of Customer Presets, Patient Database (Patient Archive, Report Archive, System
Configuration and User Defined Configuration) to a removable media.
TCP/IP Addresses, Computer Name and Software Option Password (alphanumeric string) settings
must be manually recorded.
NOTICE
Do not use this procedure when upgrading a unit’s software. Additional software and procedures, not
described here, may be needed for such upgrades.
Use this procedure if you need to reinstall the existing software on a Vivid 7.
NOTICE
In order to complete a successful restore of the Patient Database, if needed, the images must be moved
to MO disks before doing backup of Patient Database. If the backup procedure is not completed
correctly, the images and database information could be lost.