DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 3 - System Setup 3 - 89
Section 3-10
Connectivity Setup - Software v3.x
NOTE: If connected to a stand-alone network (Peer-to-Peer network with a Vivid 7 scanner, an
EchoPAC PC Workstation and an optional network printer), you should use default delivery
settings.
3-10-1 Introduction
To be able to use the network functions when connected to a hospital network, the scanner must have
a proper network address.
Before you can set up the scanner, you need to collect some information. The "Pre-quote
Worksheet for DICOM Network Information" on page 2-14 can be used for gathering this
information. Typical source for this information is the network administrator.
Follow the steps below to prepare the scanner for use on the network.
3-10-2 Contents in this Section
3-10-3 "Compatibility" on page 3-89
3-10-4 "Select TCP/IP Set-up Screen - Software Version v3.x" on page 3-90
3-10-5 "Set the Scanner’s Network Information - Software Version v3.x" on page 3-91
3-10-6 "Set the Remote Archive’s Network Information - Software Version v3.x" on page 3-92
3-10-7 "Save the New Settings" on page 3-92
3-10-8 "Connect to a DICOM Server in a Network - Software Version v3.x" on page 3-93
3-10-9 "HL7 Communication Setup - Software v3.x" on page 3-98
3-10-10 "Query/Retrieve (Q/R) Setup for Software v3.x" on page 3-102
3-10-3 Compatibility
Vivid 7 with Software Version v3.3.x (System Software v1.5.1) can communicate with:
EchoPAC PC with software version v3.3.x
EchoPAC PC with software version v4.x
EchoPAC PC with software version v5.x
EchoPAC PC with software version v6.x
EchoPAC PC with software version v7.x
Image Vault 3
Image Vault 4
EchoServer 7