8-10-1 Manpower
One person, 5 minutes
8-10-2 Tools
No tool required for installation and removal of the Filter Assembly.
8-10-3 Preparations
None
8-10-4 BEP Filter Removel Procedure
Follow these steps to remove the Filter Assembly:
1.) Press down the upper rib and pull out the upper part of the assembly before removing it completely.
2.) Lift the filter assembly away.
8-10-5 BEP Filter Installation Procedure
Follow these steps to remove the Filter Assembly:
1.) Place the two lower tags into the opening in the BEP.
2.) Push the upper part of the Filter Assembly against the BEP until the assembly is fixed by a click.
Figure 8-78 BEP Filter ASSY removal
Figure 8-79 BEP Filter ASSY installation