EasyManua.ls Logo

GE Vivid 7 - Mechanical Functions Checks

GE Vivid 7
858 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
4 - 78 Section 4-3 - Functional Checks
4-3-21 Mechanical Functions Checks
4-3-21-1 Alphanumeric Keyboard and Display Platform (Console) Checks
Figure 4-41 Console with handles
Table 4-41 Display Platform Maneuverability Checks
STEP TASK NOTE(S)
1.
Pull handle up under the platform towards you to
release horizontal lock
It is possible to maneuver the Display Platform
horizontally at present level of height
2.
Pull handle at left hand grip towards you to release the
Vertical Lock
It is possible to maneuver the Display Platform up and
down (between the vertical end stops) as long as you
pull the handle towards you
3.
Pull both handles towards you to release both Vertical
and Horizontal locks
It is possible to maneuver the Display Platform both
horizontally and vertically as long as you keep pulling
these handles towards you
Handles for horizontal and vertical lock (up
under the platform)

Table of Contents