GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
1 - 52 Section 1-5 - Labels Locations
1-5-19 Label, Internal Connections (Int.Conn.)
Located on the outside of the Front-End Card Rack’s cover (inside unit). Different versions of the label
have been used since production start.
Figure 1-7 Label on FEP2 (w/RFI)
K1 K2
K3
Cable and Connector numbering and naming system used:
IIO = Internal IO module
EIO = External IO module
BEP = Backend Processor
FEP = Frontend Processor
ACP = AC controller module
ACD = AC distribution box
ACT = AC isolation transformer module
BEPIO = IO module on the BEP
Axx : connectors on IIO with other connections than the BEP
Bxx : connectors on IIO that are connected to B EP
Cxx : connectors on BEP, except for the BEPIO
Dxx : connectors on BEPI O
Exx : connector on the BEP power supply
Fxx : connectors on AC P
Gxx : connectors on FE P
Hxx : external accesable connectors (Doppler probe)
Kxx : connectors on MODEM
Mxx : connectors on ACT
Nxx : connectors on ACD
Lxx : connectors on EIO
External IO module
69
15
M
14
69
15
M
Analogue
phone
Connect
to InSit e
M O DEM
RS-232
Connect
to InSit e
M O DEM
Service
use ONLY
(remote
power on/of f
control)
L1 L2
L3
BF-64
RFI
DC-power
supply
TX-powe r
supp ly
Board location in Front End Processor (FEP)
Monitor
Peripherals
Modem
Back End
Processor
Front End
Rack
115V
115V
or
230V
230V
230V
230V
From Isolation
transformer
230V
115V
WARNING !
Live Voltage inside,
do not open.
AC power control module
To Isolation
transformer
115V or 230V inlet
voltage selection
Voltage Selection
on Peripherals
115V = 100-1
20V
230V = 220-2
40V
(½ or 2 times i
nlet
voltag
e)
H1
G1
F1 F2 F3 F4
F5
F6
F7
F8
F9
Isolation
transformer
modul e
M1
69
15
M
230V
115V
F
an
contr
ol
81
15 9
F
A13
81
15 9
F
A12
113
14 25
M
A16
18
915
M
A15
69
15
M
A14
A12 Spare power (future option)
A13 Power to 3D box (future optio n)
A14 AC power control
A15 Signals to rotation adapter (future option)
A16 Signals to 3D box (future option)
connector
to EIO
1
25
A
F
P30
Connector to
motherboard
P31
P32
IIO seen from F
EP side
Part no. : FC314311
Rev. : 01
60,0
Cable to
C15
Modem
MT5634ZBA
Modem
power
suppl y
K4
G4
G3
AC
Distribution
bo x
N1
N2
N3
N4
N5
RX-128
Relay board
M
F
L4 L5
L6
L7
L8
L9
L10
L11 L12 L13 L14
L15
L16
L17
External IO module
(Front view)
L4: Audio out (right)
L5: Audio out (left)
L6: Trig out
L7: Foot switch
L8: Serial port
L9: Remote expose #1
L10: Remote expose #2
L11: Anal og modem
L12: Composite video output
L13: B&W video output
L14: S-Video output
L15: SVGA output
L16: USB
L17: Ethernet
RESET
TX-128
BF-64
G2
TX128 (optional)