8-15-5 Front Casters Installation Procedure
1.) Move the ball piece of the joint from old Caster to new Caster (See Figure 8-95).
2.) Align new Caster on the frame and fasten it tightly with four (4x) hexagon screws.
3.) Press the ball joint together and install the lock pin (See Figure 8-95).
4.) To replace the other Front Caster, repeat all the steps from "Front Casters Removal Procedure" on 
page 8-96, but now using the Bevel Edged board and the Wooden Wedge on the other side of the 
system.
5.) Install Bumper and Covers, see "Front Bumper Installation Procedure" on page 8-100.