8 - 90 Section 8-13 - TX Power Replacement Procedure
8-13-4 TX Power Removal Procedure
1.) Remove retaining clasp from DC Power connector.
2.) Unplug DC Power and TX Power.
3.) Unscrew two (2) screws in front of TX Power.
.
4.) Push cables aside and pull out TX Power from its socket.
8-13-5 TX Power Installation Procedure
1.) Push cables aside and install TX Power.
2.) Fasten two (2) screws.
3.) Connect DC Power Cable and secure the connector with retaining clasp.
4.) Connect HV Power Cable (TX Power Cable)
5.) Connect PCI cable to IMP2 Board.
6.) Install Right Side Cover, see "Side Covers Removal Procedure" on page 8-4.
7.) Install Front-End Cover.
Figure 8-85 DC Power, TX Power and Cabling.
HV POWER CABLE
DC POWER CABLE
TX POWER
SCREWS