GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
10 - 8 Section 10-5 - System Maintenance
10-5-2-2 Peripheral/Option Checks
If any peripherals or options are not part of the system configuration, the check can be omitted.
10-5-3 Input Power
10-5-3-1 Mains Cable Inspection
Table 10-6 GE Approved Peripheral/Hardware Option Functional Checks
Step Item Description
1DVR/VCR
Verify record/playback capabilities of the unit: Clean heads (VCR ONLY) and covers if necessary.
2 B/W Printer
Verify hardcopy output of the B/W video page printer. Clean heads and covers if necessary.
3 Color Printer
Verify hardcopy output of the Color video page printer. Clean heads and covers if necessary.
4 DICOM
Verify that DICOM is functioning properly. Send an image to a DICOM device.
5 InSite/iLinq
Verify that InSite is functioning properly. Ensure two-way remote communications.
(Warranty & Contract Customers only)
6 Footswitch
Verify that the footswitch is functioning as programed. Clean as necessary.
7ECG
Verify basic operation with customer
Table 10-7 Mains Cable Inspection
Step Item Description
1 Unplug Cord
Disconnect the mains cable from the wall and system.
2Inspect
Inspect it and its connectors for damage of any kinds.
3Verify
Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals, and that no
strands may cause a short circuit.
4 Verify
Inlet connector retainer is functional.