GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
8 - 102 Section 8-18 - Direction Lock Pedal Replacement Procedure
Section 8-18
Direction Lock Pedal Replacement Procedure
8-18-1 Manpower
One person, 15 minutes
8-18-2 Tools
Torx screwdriver T-20
8-18-3 Direction Lock Replacement Procedure
1.) Unscrew 4 screws, 3 under and 1 at the left side of the pedal.
2.) Replace the pedal with a new one.
3.) Fasten 4 screws.
Figure 8-99 Direction Lock Pedal
Direction Lock Pedal