DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7 - 7
Section 7-6
Diagnostics Tools
7-6-1 Diagnostic Procedure Summary
7-6-1-1 Purpose of this Sub-Section
These software based diagnostics tools are available on Vivid 7:
Common Service Diagnostic Interface (GSUI) and
Vivid 7’s Front-End System Test
To be able to use these Service Tools, the Back-End Processor must be up and running to some extent.
If the Back-End Processor doesn’t starts, it’s not possible to use the software service tools!
A typical approach would be to first log on to the GSUI and run the tests there. If the Acquisition test
fails, it is time to switch to Vivid 7’s Front-End System Test.
7-6-1-2 List of Service Tools Available via the Common Service Diagnostic Interface (GSUI)
Table 7-3 List of Service Tools Available via GSUI
Software Tool Description Operator
Admin-
istrator
External
Service
Page
Number
Error logs
Vivid 7 Error Logs x x x 7-11
Diagnostics
Vivid 7 Diagnostics x x x 7-26
Image Quality
[Not Populated] - - - 7-32
Calibration
DC Offset Calibration - x x 7-33
Configuration
[Not Populated] - - - 7-35
Utilities
Vivid 7 Service Utilities x x x 7-36
Replacement
[Not Populated] - - - 7-46
PM
[Not Populated] - - - 7-47