GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 86 Section 8-12 - DC Power Replacement Procedure
Section 8-12
DC Power Replacement Procedure
8-12-1 Manpower
One person, 30 minutes
8-12-2 Tools
4 mm blade screwdriver.
8-12-3 Preparations
1.) Shut down the system, see 4-2-3 "Power Shut Down" on page 4-7.
2.) Remove Right Side Cover, see "Side Covers Removal Procedure" on page 8-4.
3.) Remove the Front End’s Cover.
a.) Turn seven (7) quarter-turn screws to loosen Front-End door.
b.) Holding the door, lift it out of its retainers along the lower edge of Front-End.
DANGER
Dangerous voltages are present in this equipment. Completely power off and 
shut down the scanner before servicing.