GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
7 - 66 Section 7-8 - Acquisition Diagnostics
7-8-3 System Test
7-8-3-1 Select the Test Screen
From the Calibration screen, select TEST
or select one of the buttons (not CAL) in the lower part of
the screen. See Figure 7-70 - Select Test Window (FEP1) and Figure 7-71 - Select Test Window (FEP2).
The System Test screen (Figure 7-72/Figure 7-73) gives you access to the tests needed for
troubleshooting the Front-End Card Rack.
Figure 7-70 Select Test Window (FEP1)
Figure 7-71 Select Test Window (FEP2)
Press TEST
or press one of these
buttons
Press TEST
or press one of
these buttons