DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
4 - 32 Section 4-3 - Functional Checks
4-3-6 2D Mode (B mode) Checks
4-3-6-1 Introduction
The 2D Mode is the system’s default mode.
4-3-6-2 Preparations
Use a phantom when doing these tests.
1.) Connect one of the probes, to the scanner’s left-most active probe connector.
-See "Probe Connection" on page 3-21 for info about connecting the probes.
- See section Section 9-26 "Probes" on page 9-80 for available probes.
2.) Turn ON the scanner.
-See 4-2-2 "Power ON/ Boot UP" on page 4-3 for info about booting the scanner.
The 2D Mode window is displayed (default mode).
3.) If needed, adjust the Display’s Brightness and Contrast setting.
Figure 4-27 2D Mode Screen Example
Status Window
Control Menu