DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 4 - Functional Checks 4 - 33
4-3-6-2 Preparations (cont’d)
4.) Press 2D on the Operator Panel to access 2D Mode.
5.) These Image Controls are used to optimize the 2D picture:
-Use Gain and TGC controls to optimize the overall image together with the Power control.
-Use Depth to adjust the range to be imaged.
-Use Focus to center the focal point(s) around the region of interest.
-Use Frequency (move to higher frequencies) or Frame rate (move to lower frame rate) to
increase resolution in image.
-Use Frequency (move to lower frequency) to increase penetration.
-Use the DDP control to optimize imaging in the blood flow regions and make a cleaner, less
noisy image.
-Use Reject/Adaptive Reject controls to reduce noise in the image.
-Use the UD Clarity control to personalized appearance of the tissue. Moving the paddle to
the left creates a smoother image, though keeping boundaries sharp. Moving the paddle to
the right creates a crisper image.
Figure 4-28 Mode Dependent Keys, Zoom and Soft Menu Key
WARNINGWARNING
Always use the minimum power required to obtain acceptable images in accordance
with applicable guidelines and policies.
Depth
Gain
Mode Dependant Keys
Width
Frequency
Focus
Frame rate
Up/Down*
Left/Right*
Cineloop (in Freeze, only)
B Color Maps
Soft menu
Compress *
Reject */Adaptive Reject
1
Dynamic Range
Tilt
Contour
Diff On/Off
DDP *
Power
UD Clarity
1
2D
Freeze
Controls marked with * in the illustration are also available in Freeze and Cine Replay.
Controls marked with
1
in the illustration are used for Cardiac Probes in software version v7.0 and newer.
Zoom
Zoom
Display zoom
HR zoom