DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
7 - 64 Section 7-8 - Acquisition Diagnostics
7-8-2 Calibration Screen
After following the steps in 7-8-1-1 "Start System Test" on page 7-63, the default screen will be the
Calibration screen.
The lower part of the Calibration screen is different on units with Front-End Processor version 1 (FEP1)
and Front-End Processor version 2 (FEP2).
7-8-2-1 Functions Available from the Calibration Screen
These functions are available from the Calibration screen.
DC Offset Calibration (DC Offs. Cal) of the A/D converters in the Front-End.
Reset the calibration (Cal. Reset) to the factory default setting.
Show a graph regarding the calibration (Show graph).
Deselect/Select channels are used in manufacturing to isolate noisy channels (Channel Control).
With a linear probe connected to the system, it is possible to disconnect groups of channels either
on the TX Board or on the RX Board to isolate the noise source.
The buttons in the lower part of the screen are used for board testing (see 7-8-3 - System Test on
page 7-66).
The number of buttons in the lower row are different for units with FEP1 and FEP2. The illustration in
Figure 7-68 is the Calibration Screen on units with FEP1.
Figure 7-69 "Calibration Screen on units with FEP2" on page 7-65 illustrates the Calibration Screen on
units with FEP2.
Figure 7-68 Calibration Dialog on units with Front-End rack version - 1 (FEP1)
Channel Control
(used for factory
tests)
DC Offs. Cal
Cal. Reset
Show graph
“LEDs”
Gray - Test not done
Green - Test Passed
Red - Test failed

Table of Contents