GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
3 - 56 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-12 Remote Path
Remote Path is used for Save As, Export from Q-Analysis, and for exporting error logs with the <Alt+D>
command.
Follow this procedure to Configure Vivid 7 to use the Remote Path.
1.) Press Config (F2) and log on as ADM, see 4-2-4 "Log On to the System as ‘ADM’" on page 4-12.
2.) If not already selected, select Connectivity from the bottom row of “buttons” on the screen.
3.) Select the Tools tab.
4.) Enter the Remote Path.
Example: \\3.187.187.30\EXPORT\804
5.) Use the Check button to verify the connection.
Figure 3-51 Tools screen
REMOTE PATH
CONFIGURABLE
REMOTE PATH USER