DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 6 - Service Adjustments 6 - 3
Section 6-6
CRT Monitor Adjustments
6-6-1 Cautions and Warnings
N/A
6-6-2 Access to Adjustments
The adjustments of the monitor is done via controls (buttons) on front of the monitor.
6-6-3 Degaussing the CRT Monitor
Degaussing refers to the process of removing magnetic-field effects from the monitor. Operation of the
monitor within a magnetic field may adversely effect color purity. Degaussing can be used to correct this
problem.
To activate the manual degauss:
1.) Press either the <<< or >>> monitor adjustment button while the brightness/contrast adjustment
indicator is not displayed on the monitor. To cancel the manual degauss, press the <<< or >>>
adjustment button again.
2.) Press the toggle button for brightness and contrast.
3.) Press either the <<< or >>> monitor adjustment.
NOTE: The monitor automatically degausses at power on.