DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 8 - Replacement Procedures 8 - 7
Section 8-4
Front Cover Replacement Procedure
8-4-1 Manpower
One person, 30 minutes
8-4-2 Tools
Phillips screwdriver
8-4-3 Preparations
1.) Shut down the system, see 4-2-3 "Power Shut Down" on page 4-7.
2.) Disconnect all Probes and I/O cabling. See User manual.
3.) Remove Side Covers, see "Side Covers Removal Procedure" on page 8-4.
8-4-4 Front Cover Removal Procedure
1.) Unscrew five (5) Phillips screws that fasten it to the chassis.
2.) Pull it outwards to free it from the Top Cover and Front Bumper (See Figure 8-5).
8-4-5 Front Cover Installation Procedure
1.) Thread the same, or a new, cover so it fits in between chassis and bumper (See Figure 8-5).
2.) Thread it so it links together with Top Cover. You may have to loosen two (2) Phillips screws on the
top cover that are closest to the Front cover.
3.) Fasten Front Cover with five (5x) Phillips screws.
4.) Fasten two (2x) Phillips screws on Top Cover if these were loosened.
5.) Install Side Covers, see 8-2-5 "Side Covers Installation Procedure" on page 8-5.
6.) New Front Cover: Install correct labels, see Section 1-5 "Labels Locations" on page 1-20.
8-4-6 Verification - Functional Checks
Please perform the tests specified in Section 8-20 "Verification - Functional Checks" on page 8-105.
Figure 8-5 Front Cover Replacement
Screws
Screws