DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
8 - 38 Section 8-8 - Software Loading Procedure - Software Version v3.x
Section 8-8
Software Loading Procedure - Software Version v3.x
8-8-1 Introduction
This procedure describes how to perform a complete installation of the software on the Vivid 7 hard disk.
8-8-2 Preparations Before Software Re-installation
Export Patient Images from the internal hard disk to removable removable media .
Make backup of Customer Presets, Patient Database (Patient Archive, Report Archive, System
Configuration and User Defined Configuration) to removable media .
TCP/IP Addresses, Computer Name and Software Option Password (alphanumeric string) settings
must be manually recorded.
NOTICE
Do not use this procedure when upgrading a unit’s software. Additional software and
procedures, not described here, may be needed for such upgrades.
Use this procedure if you need to reinstall the existing software on a Vivid 7.
NOTICE
In order to complete a successful restore of the Patient Database, if needed, the images must
be moved to removable media before doing backup of Patient Database. If the backup
procedure is not completed correctly, the images and database information could be lost.