GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 39
8-8-3 Customer Provided Prerequisite
• Formatted and labelled  removable media  for Images storage
• Formatted and labelled MO Disk for Patient Archive and User Defined Settings.
8-8-4 Tools Provided With Unit or After an Upgrade
8-8-5 Preparations
1.) Switch ON the Mains Power Circuit Breaker and power up the unit, see 4-2-2 "Power ON/ Boot UP" 
on page 4-3.
2.) Store images to  removable media , see 8-8-6 "Backup - Software Version v3.x" on page 8-40.
Table 8-9   CDs with BT’03 Software, SW. v3.x
ITEM TITLE PART NUMBER DESCRIPTION
1
V7 SYSTEM SW CD, BEP1
FC200563
or 
FC200165
FC200563_02 is system software version 1.5.1 (for BEP1)
FC200165_05 is system software version 1.3.3 (for BEP1)
V7 SYSTEM SW CD, BEP2
FC200564 
or
FC200265
FC200564_02 is system software version 1.5.1 (for BEP2)
FC200265_04 is system software version 1.3.3 (for BEP2)
V7 SYSTEM SW CD, BEP2.2
FC200564
or 
FC200445
USED FOR BEPs in Vivid 7 with RFI
FC200445_01 is system software version 1.4.0 (for BEP2.2)
FC200564_02 is system software version 1.5.1 (for BEP2.2)
V7 SYSTEM SW CD, BEP4 as
BEP2 and BEP2.2 replacements
FC200825
USED for BEP4 as BEP2 and BEP2.2 replacements
FC200824_01 is system software version 1.5.x 
2
VIVID 7 & ECHOPAC PC APPL. SW
FC200567
or
FC200360
APPLICATION SOFTWARE, 
COMMON TO Vivid 7 AND ECHOPAC PC
FC200567_04 IS APPLICATION SOFTWARE VERSION v3.3.3
FC200360_07 IS APPLICATION SOFTWARE VERSION v3.1.5
VIVID 7 RFI APPL. SW FC200455
APPLICATION SOFTWARE, 
FOR Vivid 7 with RFI
FC200455_04 IS APPLICATION SOFTWARE VERSION v3.2.6