GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 4 - Functional Checks 4 - 43
4-3-8 Color Mode Checks
4-3-8-1 Introduction
Color Flow screens is 2D or M Mode screens with colors representing blood or tissue movement.
Color Flow may be selected both from 2D mode or from M mode or a combination of these.
4-3-8-2 Preparations
Use a phantom when doing these tests.
1.) Connect one of the probes, to the scanner’s left-most probe connector.
-See "Probe Connection" on page 3-21 for info about connecting the probes.
-See Chapter 9 - , section 9-26 for available probes.
2.) Turn ON the scanner.
-See 4-2-2 - Power ON/ Boot UP on page 4-3 for info about booting the scanner
The 2D Mode window is displayed (default mode).
4-3-8-3 Color 2D Mode
Press COLOR
.
4-3-8-4 Color M Mode
1.) Press MM
 (M Mode) to bring up an M Mode display on the screen.
2.) Press COLOR
.