DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
3 - 58 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-14 Set Up Connection to a DICOM Server in a Network
In this case the Vivid 7 is configured to work with DICOM servers in a network environment. Images are
first saved on the local image buffer on the scanner. At the end of the examination the images are sent
to the DICOM server via a DICOM spooler and to the local database, depending on dataflows.
This scenario requires that the scanner is configured to be connected to DICOM servers as described
below.
3-9-14-1 Overview
To connect to the DICOM server, the following information has to be entered in the scanner:
The DICOM server IP address.
The DICOM server port number.
The DICOM server AE title (the server application’s name).
Name of Device.
3-9-14-2 DICOM Server IP Address Setting on the Scanner
1.) Press CONFIG
(F2) and log on as adm, see 4-2-4 "Log On to the System as ‘ADM’" on page 4-12.
2.) Select CONNECTIVITY
(in the lower part of the window).
3.) Select the DATAFLOW
tab.
4.) Select the arrow to the right for the Name field to list all dataflows in a pull-down menu, see Figure
3-53.
5.) Select the dataflow you want to configure from the Dataflow Name pull-down list, see Figure 3-53.
Figure 3-53 Select Dataflow
ARROW
SELECT DATAFLOW
FROM DATAFLOW NAME
PULL-DOWN MENU.