EasyManua.ls Logo

GE Vivid 7 - ECG Check

GE Vivid 7
858 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
4 - 68 Section 4-3 - Functional Checks
4-3-17 ECG Check
4-3-17-1 Introduction
The ECG capability on this unit, is intended as use as a trigger for measurements, but can also be
viewed on the screen.
4-3-17-2 Parts needed
ECG Harness, P/N:16L0026 + P/N:16L0028
ECG Pads, (3 pc)
or
ECG simulator
4-3-17-3 Preparations
None
4-3-17-4 ECG Check
Table 4-32 ECG Check
Step Task Expected Result(s)
1.
Connect the ECG harness to the connector on the front
of the system
The unit displays a straight curve along the bottom
edge of the image sector on the screen.
2.
Connect the three leads to a ECG simulator,
or
Fasten the three ECG Pads to your body and connect
the three leads to respective ECG Pad
- When connecting, the signal on the screen will be
noisy
- When the connection is completed, a typical clean
ECG signal is displayed.

Table of Contents

Related product manuals