DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 5 - Components and Functions (Theory) 5 - 61
5-3-21 Radio Frequency Interface Board, RFI
5-3-21-1 General Description
The RFI board was introduced into the manufacturing in the 2nd half of 2003 together with the Front-
End Processor version 2 (FEP2).
The functionality of the following boards are either implemented on the RFI board or on the BEP:
Front-End Controller board
RF & Tissue board
Spectrum Doppler Processor board (some filtering done on RFI, the rest is done by BEP).
Image Port board.
Tasks Done by the RFI board:
System Master Clock generator
Transmitter Control
ATGC Control
Receive Focusing Control
Analog Test Signal Generator
High Voltage Control and supervision
Probe Sensing
Control of Relay Board
Doppler Audio D/A conversion
Temperature Sensing for TEE probes
Step Motor Control for TEE probes
Temperature Sensing for circulated air
5-3-21-2 Signal Flow
The digitized signals from the Beam Formers, are connected to the RFI module. The Digital Signal
Processor on the RFI board together with software on the BEP, performs the adequate signal
conditioning for the different data types; Tissue, Doppler and Color Flow.
In addition, Doppler Audio Processing is done by the RFI board. The Doppler Spectrum Processing is
done by the BEP.
The RFI extracts Tissue Data, Spectrum Data and IQ & Color Data from the digital data stream on its
input and it also performs RF filtering and different types of Tissue Processing.
5-3-21-3 RF Processing
Different types of RF processing is performed depending on later usage of the data;
RF Demodulation
Digital Time Gain Compensation
Filtering
Decimation (data that are not needed for the processing, is removed)