DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 3 - System Setup 3 - 55
3-9-10 Set the Remote Archive’s Network Information
To be able to connect to a remote archive, on a remote computer or server, you must configure Vivid 7
to communicate with it.
In the Remote Archive Setup area of the Tcpip screen, see example in Figure 3-48, enter the;
1.) Remote Archive IP address. (Default IP Address from factory: 10.0.0.4).
2.) Remote Archive Name. (Default Remote Archive Name from factory: ECHOPAC7-000001).
.
3-9-11 Save the New Settings
1.) Select SAVE SETTINGS to save the new settings.
.
A Warning is displayed on the screen, see illustration below.
2.) The new settings are saved to a common settings file. After a restart, the settings are also included
in other screens.
3.) Restart Vivid 7 to activate the changes.
Figure 3-48 TCP/IP Set-up for Vivid 7
Figure 3-49 Save New TCP/IP settings
Figure 3-50 Warning
IP ADDRESS FOR THE
REMOTE ARCHIVE
COMPUTER NAME FOR
THE REMOTE ARCHIVE
SAVE SETTINGS