GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 3 - System Setup 3 - 133
Section 3-12
Ethernet Switch / Hub
An Ethernet Switch is used to connect Vivid 7 to an EchoPAC PC and a network printer in a Peer-to-
Peer network. It can also be connected to the hospital’s network for access to a DICOM server.
The Ethernet Switch described here is included as an example. Both five port and eight port switches
have been delivered with EchoPAC PC. Please read and follow the product documentation for the
actual switch before installing/configuring.
:
Figure 3-143 Ethernet Switch Connections (8-port example)
Figure 3-144 Ethernet Switch (5-port example)
Table 3-22 MDI Button Position Descriptions
POSITION LABEL FUNCTION
X
To PC
To connect an end station or a Server to the Port
=
To HUB
To connect a HUB or another Switch to the Port
Vivid i Vivid 3
Vivid 4
ECHOPAC
PC
Vivid 7
ECHOPAC
PC
NETWORK
PRINTER
OPTIONAL CONNECTION
TO HOSPITAL’s ETHERNET
LEDS
MDI Button controls the
fifth connector, making it
crossed or non-crossed.
Status LEDs
Ethernet Cable
connectors