DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
3 - 132 Section 3-11 - Connectivity Setup - Software v2.x
3-11-10-9 Performing a Study
1.) After restart, hit New Exam (you’ll see that DICOM Server is the default dataflow), enter a last name
and patient ID and start scanning and image acquisition. Images are stored in the local image buffer
on the system.
2.) After having finished the study, Select New Exam and select All when asked if you want to save.
The images will now be transferred to the server via a spooler.
3.) To view the status of the spooler job, press the F4
key on the alphanumeric keyboard. You can
select the Refresh button to get an update of the status of the spool job.
4.) Once the jobs are transferred, they will be deleted from the spooler.
3-11-10-10 Going Offline
1.) If you go offline with the scanner, you can still use the DICOM Server dataflow, and images will be
saved into the spooler.
2.) Then, after connecting to the network, go into the spooler and press RESEND
. All selected studies
will then be sent to the server.
NOTE: If the Check button fails directly, the AE Title is probably wrong.
If the Check button fails after a long time (corresponding to the timeout), the IP address or Port
Number is probably wrong.