DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 3 - System Setup 3 - 131
3-11-10-8 Setting the DICOM Server as Default Dataflow - Software v2.x
1.) Finish the verification in "Check the Connection to the DICOM Server (Ping) - Software v2.x" on
page 3-130.
You need to have green check marks on both Hardware Ping and DICOM Ping before you continue.
1.) Select the DATAFLOW
tab.
2.) Select “DICOM Server” dataflow on the “Dataflow Name” drop-down menu.
.
3.) Check the Default box.
.
4.) Exit Config and restart the scanner (by hitting CTRL/ALT/R at the same time).
Figure 3-141 Selecting DICOM Dataflow Services
Figure 3-142 Set DICOM as Default Dataflow
DATAFLOW tab
Dataflow Name
drop-down menu
Select
“DICOM Server”
Select this check box to
make DICOM the Default
dataflow