DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 1 - Introduction 1 - 19
1-4-5 Electrical Safety
1-4-5-1 Safe Practices
Follow these guidelines to minimize shock hazards whenever you are using the scanner;
The equipment chassis must be connected to an electrical ground.
The unit is equipped with a three-conductor AC power cable. This must be plugged into an approved
electrical outlet with safety ground.
The power outlet used for this equipment should not be shared with other types of equipment.
Both the system power cable and the power connector must meet international electrical standards.
1-4-5-2 Probes
Follow these guidelines before connecting a probe to the scanner;
Inspect the probe prior to each use for damage or degradation to the;
- housing
- cable strain relief
-lens
- seal
Do not use a damaged or defective probe.
Never immerse the probe connector or adapter into any liquid.
WARNINGWARNING
CONNECTING A Vivid 7 SCANNER TO THE WRONG VOLTAGE LEVEL WILL MOST
LIKELY DESTROY IT.