GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 8 - Replacement Procedures 8 - 87
8-12-4 DC Power Removal Procedure
1.) Unscrew three (3) screws as shown in Figure 8-82.
2.) Unplug PCI cable from the socket on IMP2 board.
3.) Pull the DC Power a little to get access to power cables.
4.) Remove retaining clasp and unplug Power Cable and HV Cable.
5.) Remove DC Power.
Figure 8-82 DC Power and Cabling.
Screws
HV power cable
DC power cable
PCI cable
IMP2 Board
DC Power