GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 4 - Functional Checks 4 - 77
4-3-20-7 Check Cineloop from Video
Prerequisite
A Video Tape in the VCR with at least one recorded examination.
Procedure
4-3-20-8 Turn OFF Power to Vivid 7
See 4-2-3 "Power Shut Down" on page 4-7.
Table 4-40 VCR Cineloop Checks
Step Task Expected Result(s)
1.
When in video play back (of an examination), press
FREEZE
.
The last part of the video is stored in the computer’s
memory as a cine loop.
2.
Use the Trackball to move back and forth in the cine
loop.
The Cineloop is played back according to your
manipulating of the trackball.