DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 10 - Care & Maintenance 10 - 15
10-7-4-1 Meter Procedure
Follow these steps to test the ground wire resistance.
1.) Turn the Vivid 7 unit OFF.
2.) Plug the unit into the meter, and the meter into the tested AC wall outlet.
3.) Plug the black chassis cable into the meter's “CHASSIS” connector and attach the black chassis
cable clamp to an exposed metal part of the Vivid 7 unit.
4.) Set the meter's “FUNCTION” switch to the RESISTANCE position.
5.) Set the meter's “POLARITY” switch to the OFF (center) position.
6.) Measure and record the ground wire resistance.