EasyManua.ls Logo

GE Vivid 7 - Page 445

GE Vivid 7
858 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 5 - Components and Functions (Theory) 5 - 99
5-4-10 UPS Battery Description (cont’d)
The UPS batteries will drain slightly when the system is in the off condition. The off condition is defined
as:
Circuit Breaker in the OFF position
System unplugged from the wall outlet
If the system is left in the OFF condition for an extended period of time (3 to 5 days or more), the system
may not boot up or may beep when turned ON. Should this occur, the system needs to recharge the
UPS batteries. This could take 15 minutes to as long as 10 hours uninterrupted, depending on the
battery age, system input voltage and system temperature. To eliminate draining the UPS batteries, the
system should remain plugged into the wall outlet with the circuit breaker in the ON position.

Table of Contents

Related product manuals