GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 3 - System Setup 3 - 7
3-3-4 Receiving the Vivid 7
3-3-4-1 Examine All Packages
Examine all packages closely at time of delivery, as described in the procedure below.
Table 3-5 Examine All Packages
STEP TASK ILLUSTRATIONS
1
Is damage apparent?
• If yes; continue with the instructions in subsection
3-3-4-2 - Damage in Transportation.
• If no; continue with the next step.
2 Is the Shock Indicator red colored inside the middle of the
indicator?
• If yes: The Shock Indicator has been activated.
Continue with the instructions in subsection
3-3-4-2 - Damage in Transportation.
• If no: Continue with step 2.
3 Is the Tilt Indicator red colored inside the middle of the
indicator?
• If yes: The Tilt Indicator has been activated.
Continue with the instructions in subsection
3-3-4-2 - Damage in Transportation before you continue
with the next step.
• If no: Continue with the instructions in 3-3-5 - Unpacking
Vivid 7.
4 The two upper hinges on the Front Side and the Rear Side of
the wooden transportation bow have been sealed with red
plastic seals, marked GE Vingmed Ultrasound and a serial
number. Verify that the four red plastic seals are intact at
arrival.
If seals are broken:
If broken, it indicates that the container may have been
opened after it left the manufacturer.
Continue with the instructions in subsection
3-3-4-2 - Damage in Transportation.
Continue with the instructions in 3-3-5 - Unpacking Vivid 7.
RED COLOR
RED COLOR