3-6-2-2 Electrical Specifications for Vivid 7
Table 3-10 Electrical Specifications for Vivid 7 sheet 1 of 3
GE VINGMED
PART NUMBER DESCRIPTION VOLTAGE TOLERANCES CURRENT FREQUENCY
FD000010
VIVID 7 DIMENSION CONSOLE,
LCD MONITOR
230 VAC ±10% 5 A 50-60 Hz
FD000020
VIVID 7 DIMENSION CONSOLE,
LCD MONITOR
100-120 VAC ±10% 10 A 50-60 Hz
FD000120
VIVID 7 PRO CONSOLE,
LCD MONITOR
230 VAC ±10% 5 A 50-60 Hz
FD000130
VIVID 7 PRO CONSOLE,
LCD MONITOR
100-120 VAC ±10% 10 A 50-60 Hz
FD000140
VIVID 7 DIM. W/ MULTI-DIM. &
VOLUME,
LCD MONITOR
230 VAC ±10% 5 A 50-60 Hz
FD000150
VIVID 7 DIM. W/ MULTI-DIM. &
VOLUME,
LCD MONITOR
100-120 VAC ±10% 10 A 50-60 Hz
FD000160
VIVID 7 DIM. W/ MULTI-DIM.
IMAGING,
LCD MONITOR
230 VAC ±10% 5 A 50-60 Hz
FD000170
VIVID 7 DIM. W/ MULTI-DIM.
IMAGING,
LCD MONITOR
100-120 VAC ±10% 10 A 50-60 Hz
FD000180
VIVID 7 DIM. W/ VOLUME
ULTRASOUND,
LCD MONITOR
230 VAC ±10% 5 A 50-60 Hz
FD000190
VIVID 7 DIM. W/ VOLUME
ULTRASOUND,
LCD MONITOR
100-120 VAC ±10% 10 A 50-60 Hz
FC000890
VIVID 7 DIMENSION
MULTIDIMENSIONAL & Volume
(BT’06) 230 VAC
230 VAC ±10% 5 A 50-60 Hz
FC000900
VIVID 7 DIMENSION
MULTIDIMENSIONAL & Volume
(BT’06) 120 VAC
110-120 VAC ±10% 10 A 50-60 Hz
FC000910
VIVID 7 DIMENSION (BT’06) 230 VAC
230 VAC ±10% 5 A 50-60 Hz
FC000920
VIVID 7 DIMENSION (BT’06)
100-120 VAC
110-120 VAC ±10% 10 A 50-60 Hz
FC000930
VIVID 7 PRO (BT’06) 230 VAC
230 VAC ±10% 5 A 50-60 Hz
FC000940
VIVID 7 PRO (BT’06) 100-120 VAC
110-120 VAC ±10% 10 A 50-60 Hz
FC000950
VIVID 7 DIMENSION
MULTIDIMENSIONAL IMAGING
(BT’06) 230 VAC
230 VAC ±10% 5 A 50-60 Hz
FC000960
VIVID 7 DIMENSION
MULTIDIMENSIONAL IMAGING
(BT’06) 120 VAC
110-120 VAC ±10% 10 A 50-60 Hz