DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
3 - 76 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-16-2 Create a New Dataflow with Worklist, Database and DICOM Storage (cont’d)
3.) Add the DICOM Storage Output, highlight it, select properties, and name it (i.e.: DR PACS Store).
Go to the IP Address field, scroll down and select <Modify>. This will allow you to edit or add IP
Addresses (New Destinations/devices).
If the correct IP Address is already in the list, just select it.
If you create the Dataflow in any other order or remove and add an entry later, you may or will receive
“you are not connected to the Database” errors when trying to create a patient locally. The Worklist will
work, however they can NOT save patients “created” on the Vivid 7 to the Archive while using that
dataflow.
You can change the information on each entry in a Dataflow as long as you do NOT remove or add to
the Dataflow out of sequence.
You should also set Direct Search to “All patients” for the new dataflow.