DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 4 - Functional Checks 4 - 11
4-2-3-1 Standby Mode (cont’d)
NOTE: Connect the power cable to the mains power outlet again as soon as possible and at least
before you switch on the Mains Circuit Breaker and press the ON/STANDBY
key.
NOTE: When turning ON a system from standby mode, it takes a few seconds before it responds. Do not push
the ON/STANDBY
button again during this period. A second push will initiate a full shutdown.
4-2-3-2 Complete Power Down
NOTE: This function is not available from Standby mode.
Complete Power Down Procedure
1.) Press the ON/STANDBY
key on the Operator Panel once to display the System - Exit menu.
2.) Select Shutdown to do a complete power down of the unit.
Back-End Processor will first turn OFF the Scanner activity and print the message “Please wait -
Shutdown in progress” in the LCD display on the Operator Panel. Then it starts to shut down itself.
The time to turn down the unit including the Back-End Processor, may vary from 10 seconds up to
approximately 1 minute. The last thing that shuts down, is the light in the LCD displays, indicating
that you can continue with the next step.
3.) Switch OFF the mains power CIRCUIT BREAKER
(located at the rear of the unit). This will cut
power distribution within the unit.
If the Back-End Processor is still running when you switch off the mains power CIRCUIT
BREAKER, the internal Un-interrupted Power Supply (UPS) will switch over to battery power and
supply the Back-End Processor with power until it has shut down properly.
A periodic high-pitched sound alarm indicates that the Back-End Processor is running on battery
power.
.
Figure 4-9 Circuit Breaker located on rear of the unit
CIRCUIT BREAKER